SE-12 Series Electrocardiograph User Manual Basic Operation Guidance
- 39 -
4.3 Work Mode Description
There are five work modes in SE-12 series electrocardiograph.
In the auto mode, the ECG data can be analyzed, saved, printed and
transmitted. The lead groups are switched automatically according to the lead
sequence during the printing course. After the ECG waves of one lead group
are printed within a certain time, the system switches to print ECG waves of
another lead group automatically. 1mV calibration marks will be printed at the
beginning of an ECG report.
In the manual mode, you can determine the lead group to be displayed and
printed. Pressing the Left or Right arrow can switch among the lead groups.
In the rhythm mode, the ECG data can be saved and transmitted. You can print
60s rhythm-lead ECG waveform of one lead in the Single Lead style or 20s
rhythm-lead ECG waveform of three leads in the Three Leads style.
In the R-R analysis mode, you can select a lead to print its R-R histogram, R-R
trend chart, 180s compressed ECG waveform and all the R-R interval values.
Only configurable for SE-12 Express/SE-1200 Express/SE-1201
In the VCG mode, X, Y, Z waves and frontal, horizontal, sagittal planes can be
displayed. 10s sampled ECG data can be analyzed, and vector waves, vector
loops, measure information, diagnosis information can be printed.
.