SE-12 Series Electrocardiograph User Manual System Setup
- 50 -
Chapter 10 System Setup
Select Setup on the main screen1 to display the System Setup screen.
10.1 Work Mode Setup
Items
Default
Default (Only in the U.S.)
Mode Options
Auto, Manual, Rhythm
Auto, Manual, Rhythm, R-R
Display Style
2
4+1R
Rhythm Style
Three Leads
Three Leads
Sampling Mode
Real-time Sample
Real-time Sample
Duration
(Periodic Sample)
60 min
60 min
Interval
(Periodic Sample)
1 min
1 min
Preview
Off
Off
Auto Arrhythmia Detection
Off
Off
10.2 Filter Setup
Items
Default
Default (Only in the U.S.)
AC filter
On
On
EMG filter
Off
Off
DFT filter
0.67Hz
0.67Hz
Lowpass filter
100Hz
100Hz
NOTE:
1. AC frequency can be set to 50Hz or 60Hz on the Advanced Setup screen according
to local mains supply specifications.
2. Only when EMG Filter is set to Off, can the setting of Lowpass Filter be effective.
3. To pass the distortion test, the electrocardiograph has to be configured with the
highest bandwidth in filter settings. Otherwise, ECG signal may be distorted.