STP-L301/L451 Series Instruction Manual
8-11
Operation
Table 8.8 shows rated contacts for relays CR4, 5, 6, 7, 8, 9 and 10 (see Figure 8.4).
Table 8.8 Rated Contacts for Relays CR4, 5, 6, 7, 8, 9 and 10
Resistance Load
(COS
φ = 1)
Induction Load
(COSφ = 0.4)
(L/R = 7 ms)
125 Vac 0.4 A 125 Vac 0.2 A
Rated Load
30 Vdc 2 A 30 Vdc 1 A
Rated Current 3 A
Maximum Contact
Point Voltage
250 Vac
220 Vdc
Maximum Contact
Point Current
3 A
Maximum Open/Close
Capacity
AC: 50 VA
DC: 60 W
AC: 25 VA
DC: 30 W
Minimum
Applicable Load
10 mVdc, 10 uA
8.3.3 Remote Cables (Optional Accessories)
Remote cables for CON4 are provided as optional accessories.
Remote cables are available in two types: single-side connector type and both-side connector
type. Contact Edwards for details.
Table 8.9 shows connections of remote cables of single-side connector type.
Table 8.9 Remote Cable Connections (Single-Side Connector Type)
Connector
Pin No.
Cable Color
Connector
Pin No.
Cable Color
1 Brown 13 White/Orange
2 Red 14 White/Yellow
3 Orange 15 White/Green
4 Yellow 16 White/Blue
5 Blue 17 Green
6 Purple 18 White/Black/Brown
7 Gray 19 White/Black/Red
8 White 20 White/Black/Orange
9 White/Black 21 White/Black/Yellow
10 White/Brown 22 White/Black/Green
11 White/Red 23 White/Black/Blue
12, hood Shielded 24 White/Purple, White/Gray