Be careful to use this function, very professional
skill is needed, otherwise has risk of file broken.
Auto Start
ON
E-connect S integrated apex locator, if the lip hook
is connecting with patient’s lip, when the endo file
entering root canal, the motor will start
automatically.
Press < or > to shut off this function if not
expected, press to start and stop the motor.
WARNING
The motor will start automatically if handpiece
(without insulating sleeve) or file touch the
patient’s lip or operator’s fingers (without
insulating glove), take care to avoid this, the file
rotated by motor has risk of injure someone.
Auto Stop
OFF
When the endo file out of root canal, the motor will
not auto stop with default setting, Press <
or > to select auto stop “ONif needed.
Apical Action
Reverse
Because of integrated apex locator, when the file
reaches the reference point, the motor will
response according to setting, it can be Reverse,
SlowDown, Stop and Off.
Press < or > to change.
Reverse: rotation direction changing till the file
upward a little bit by operator, rotation direction will
change back again.
SlowDown: rotation slowdown when approach the
reference point, will reverse if reach.
Stop: rotation stop when reach the reference point,
upward a little bit and will rotate again.
Off: rotating as usual even if reach the reference
point.