Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
field should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The E-connect S is intended for use in the electromagnetic environment specified below.
The customer or the user of the E-connect S should assure that it is used in such an
environment.
Immuni
ty test
IEC
60601 test
level
Comp
liance
level
Electromagnetic environment - guidance
Conduct
ed RF
3 Vrms
150 kHz
to 80 MHz
3 V
Portable and mobile RF communications equipment
should be used no closer to any part of the E-connect S,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
1.2dP=
1.2dP=
80 MHz800 MHz
2.3dP=
800 MHz2.5 GHz
IEC
61000-4-
6
V/m
Radiated
RF