ELIS PLZEŇ a. s., Luční 425/15, 301 00 Plzeň, Czech Republic, Phone: +420/377 517 711, Fax: +420/377 517 722 Es90420K/c
Warning:
If not agreed otherwise with the customer, flowmeters are calibrated with the 0 to 1 kHz frequency output.
Upon agreement with the manufacturer, flowmeters can be calibrated by means of other outputs, i.e. pulse
or current.
Guaranteed parameters of measurement accuracy always apply only to the electrical output used for the
flowmeter calibration. The other outputs, not use for calibration purposes, are recommended to be used only
for measurements with the permissible accuracy that is 1 to 2 % lower than the accuracy achieved when
calibrated with the frequency output.
During a control measurement or metrological verification carried out by the customer it is necessary to use
the same electrical output that was used for calibration by manufacturer.
12. PRODUCT ORDERING
When ordering the FLONET FN20xx.1 electromagnetic flowmeters, use a specific order number to be
determined based on the following table.