Troubleshooting (continued)Table 2-1:
Error Recommended action(s)
Electronics Temperature is Out Of Nominal
Range
Temperature of the electronics is out of
nominal operating range (below -40 °C or
above 100 °C) which could lead to a system
failure.
Attempt to warm or cool the meter elec-
tronics housing.
If the electronics is mounted to the meter
and the process fluid in the meter is over
60 °C, you must remote mount the electron-
ics off of the meter body.
Collect a Maintenance log using Daniel Me-
terLink while the meter is experiencing the
issue and contact your Daniel service repre-
sentative.