Maintenance (continued)Table 2-2:
Daniel MeterLink utility Diagnostics Action(s)
the transducer face, or path
length configured incor-
rectly).
• Adjust SSMin or SSMax only
if other checks pass (con-
sult a Daniel Service repre-
sentative before making
these adjustments).
Meter Monitor (Detailed) view
Meter Data List
• Electronics Temperature of
out range
• Temperature of the elec-
tronics is out of nominal
operating range below
-40 °C or above 100 °C
(-40 °F or above 212 °F).
- Heat or cool the meter
electronics housing. If
operating temperature
exceeds 60oC, remote
mount the Transmitter
Electronics Enclosure.
Meter Monitor (Detailed) view
Meter Data List
• Electronics voltage out of
range
• Systems voltages are valid
if 1.0V, 1.2V, 2.5V, 3.3V or
the Acquisition Module val-
id voltages are 1.2V, 2.5V
or 3.3V.
• Replace the CPU Module if
one or more of the System
Voltages is out of range.
• Replace the Acquisition
Module if one or more of
the voltages is out of range.
Troubleshooting
Maintenance and Troubleshooting manual 31