Transmitter Electronics Enclosure and Base Enclosure security seal
removal
Figure 3-25:
A. Transmitter electronics enclosure
B. Base enclosure bolts and security seals
4.
Use a 6 mm Allen wrench and remove the two hex head bolts securing the
Transmitter Electronics Enclosure to the Base Enclosure.
Meter repairs
82
Gas Ultrasonic Flow Meters
91
93
Table of Contents
Default Chapter
9
Table of Contents
9
Chapter 1 Maintenance
11
Precautions for Meter Maintenance
11
Field Hydrostatic Pressure Testing Procedures
12
T-Slot Transducer Assembly and Mount
13
Routine Maintenance
14
Maintenance Logs and Reports
14
Pipeline Cleaning Maintenance
18
Visual Inspection
18
Chapter 2 Troubleshooting
21
Meter Status Alarms
21
Check Status
22
System Alarm
23
Chord A, Chord B, Chord C and Chord D Alarm
23
Field I/O Alarm
23
Validity Alarm
24
Comms Alarm
24
Communications
24
Troubleshooting the Meter
24
Meter Maintenance
36
Unable to Connect Direct Serial or External Serial Modem
45
Unable to Connect to Meter
46
Ethernet Connections
46
Direct Serial Connections
46
Troubleshoot Maintenance Log Files and Trend Files
46
Files Do Not Appear in Workbook
46
Microsoft ® Excel ® Log/Export Options Are Not Available
47
Maintenance Logs or Trend Files Are Not Created
47
Chapter 3 Meter Repairs
51
Prior Precautions
51
Field Hydrostatic Pressure Testing Procedures
53
T-Slot Transducer Assembly and Mount
54
T-Slot Transducer Removal and Installation Procedures
55
Transducer Removal with Extractor Tool
56
Transducer Removal Without Extractor Tool
58
Transducers Installation Procedure
61
Replace Transformers Procedure
63
Modifying the Calibration Parameters
64
Transducer Holder Removal and Installation
66
Remove the Transducer Holder with the Extractor Tool
67
Remove the Transducer Holder Without the Extractor Tool
67
Install the Transducer Holder with the Extractor Tool
69
Install the Transducer Holder Without the Extractor Tool
69
Transducer Cable Removal and Installation
72
Remove Transducer Cables
73
Install Transducer Cables
74
Replace the Meter Electronics
78
Replace the CPU Module or Optional I/O Module
79
Replace the Fuse
82
Replace Backplane, I.S. Barrier or Power Supply Board
83
Replace the Adquisition Module
90
Appendix A Conversion Factors
97
Conversion Factors Per Units of Measurement
97
K-Factor and Inverse K-Factor
98
Appendix B Engineering Drawings
101
Daniel 3410 Series Ultrasonic Flow Meter Drawings
101
Other manuals for Emerson Daniel 3410 Series
Operation Manual
192 pages
Instructions For Use, Maintenance And Installation
110 pages
Installation Manual
140 pages
5
Based on 1 rating
Ask a question
Give review
Questions and Answers:
Need help?
Do you have a question about the Emerson Daniel 3410 Series and is the answer not in the manual?
Ask a question
Emerson Daniel 3410 Series Specifications
General
Brand
Emerson
Model
Daniel 3410 Series
Category
Measuring Instruments
Language
English
Related product manuals
Daniel Ultrasonic 3410 Series
94 pages
Emerson Daniel 3415
172 pages
Emerson Daniel 3414
192 pages
Emerson Daniel Series
106 pages
Emerson daniel 1500 series
114 pages
Daniel Senior Orifice Fitting Series
138 pages
Emerson 3410 Series
96 pages
Emerson Dixell XWEB300D PRO
20 pages
Micro Motion 7829 Viscomaster Dynamic
12 pages
Emerson AMS
78 pages
Emerson CSI 2140
404 pages
Emerson FloBoss 103
106 pages