9.1.3 Configure burst communications
Burst mode is a mode of communication during which the transmitter regularly broadcasts HART digital
information to the network via the primary mA output.
Configure HART burst messages
Display
Menu > Configuration > Inputs/Outputs > Channel A > HART Settings > Burst Message x
ProLink III Device Tools > Configuration > Communications > Communications (HART) > Burst Mode
Field Communicator Configure > Manual Setup > HART > Burst Mode
Configuration and Use Manual Configure digital communications
MMI-20048166 April 2019
Configuration and Use Manual 111