6.4.1 Default settings for totalizers and inventories
Totalizer or
inventory
Source (process variable
assignment)
Direction Name of totalizer
Name of inventory
1 Mass flow Forward Only Mass Fwd Total
Mass Fwd Inv
2 Volume flow Forward Only Volume Fwd Total
Volume Fwd Inv
3 Temperature-corrected volume
flow
Forward Only API Volume Fwd Total
API Volume Fwd Inv
4 Gas standard volume flow Forward Only GSV Fwd Total
GSV Fwd Inv
5 Standard volume flow Forward Only Standard Vol Fwd Total
Standard Vol Fwd Inv
6 Net mass flow Forward Only Net Mass Fwd Total
Net Mass Fwd Inv
7 Net volume flow Forward Only Net Vol Fwd Total
Net Vol Fwd Inv
6.5 Configure logging for totalizers and inventories
Display
Not available
ProLink III Device Tools > Configuration > Totalizer Log
Field Communicator Not available
The transmitter can write the current value of four totalizers or inventories to a log, at user-specified intervals.
You can generate a log file from this data for viewing and analysis.
Procedure
1. Specify the date on which totalizer logging will begin.
You must specify a future date. If you try to specify the current date, the transmitter will reject the
setting.
2. Specify the time at which totalizer logging will begin.
3. Specify the number of hours between records.
4. Select up to four totalizers or inventories to be logged.
6.6 Configure Process Variable Fault Action
Display
Menu > Configuration > Alert Setup > Output Fault Actions
ProLink III Device Tools > Configuration > Fault Processing
Field Communicator Configure > Alert Setup > Output Fault Actions > Process Var Fault Action
Configure advanced options for process measurement Configuration and Use Manual
April 2019 MMI-20048166
76 Micro Motion 4200 Transmitters

Table of Contents