EasyManua.ls Logo

Emerson Micro Motion 4200 - Transmitter Does Not Communicate; API Referral Problems

Emerson Micro Motion 4200
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15.2.52 Watchdog Error
Alert
Electronics Failed
Cause
The watchdog timer has expired.
Recommended actions
1. Reboot or power-cycle the transmitter to see if the alert clears.
2. If the alert persists, replace the transmitter.
3. Contact customer support.
15.3 Transmitter does not communicate
If the transmitter does not appear to be communicating, the wiring may be faulty or the communications
device may be incompatible.
For HART network communications, perform the procedures in later topic to be linked.
For communication using a communication device, check the wiring and the communication device. Refer
to either Using ProLink III with the transmitter or Using a Field Communicator with the transmitter.
If you can read data from the transmitter but cannot write data (that is, you cannot start, stop or reset
totalizers or change transmitter configuration), check to see if the transmitter is write-protected, and disable
write-protection if required.
15.4 API referral problems
15.4.1 Extrapolation alert is active
Cause
Line pressure, line temperature, or line density is outside the range of the configured API table.
Recommended actions
1. Check your process conditions against the values reported by the device.
2. Verify the configuration of the API referral application and related parameters.
15.4.2 Inaccurate referred density reading
Cause
Inaccurate density measurement
Inaccurate temperature measurement
Incorrect reference conditions
Incorrect API table selection
Troubleshooting Configuration and Use Manual
April 2019 MMI-20048166
174 Micro Motion 4200 Transmitters

Table of Contents

Other manuals for Emerson Micro Motion 4200

Related product manuals