12. Check for two-phase flow.
13. Check for radio frequency interference.
14. Contact customer service.
15.8.5 Inaccurate flow rate
Cause
Wiring problem
Inappropriate measurement unit
Incorrect flow calibration factor
Incorrect meter factor
Incorrect density calibration factors
Incorrect grounding
Two-phase flow
Problem with receiving device
Incorrect sensor zero
Recommended actions
1. Check the wiring between the sensor and the transmitter.
2. Verify that the measurement units are configured correctly for your application.
3. Verify all of the characterization or calibration parameters. See the sensor tag or the calibration sheet
for your meter.
4. Zero the meter.
5. Check the grounding of all components.
6. Check for two-phase flow.
7. Verify the receiving device, and the wiring between the transmitter and the receiving device.
8. Check the sensor coils for electrical shorts. If you find problems, replace the sensor
9. Replace the core processor or transmitter.
15.9 Frequency Output problems
15.9.1 No FO
Cause
Stopped totalizer
Process condition below cutoff
Fault condition if Fault Action is set to Internal Zero or Downscale
Two-phase flow
Flow in reverse direction from configured flow direction parameter
Configuration and Use Manual Troubleshooting