11.3 Read totalizer and inventory values
Display Menu > Operations > Totalizers > See Totals
ProLink III Device Tools > Totalizer Control > Totalizers
Device Tools > Totalizer Control > Inventories
Field Communicator Overview > Totalizer Control
Totalizers keep track of the total amount of mass or volume measured by the transmitter since the last
totalizer reset. Inventories keep track of the total amount of mass or volume measured by the transmitter
since the last inventory reset.
11.4 Start, stop, and reset totalizers and inventories
Related information
Start, stop, and reset totalizers using the display
Start, stop, and reset totalizers using ProLink III
11.4.1 Start, stop, and reset totalizers using the display
You can start and stop each totalizer or inventory independently. You can start and stop all totalizers and
inventories as a group. You can reset each totalizer or inventory independently. You can reset all totalizers as
a group.
When a totalizer or inventory is started, its value increases or decreases depending on the interaction of the
flow direction parameters. It continues tracking flow until it is stopped.
When a totalizer or inventory is reset, its value is set to 0. You can reset a totalizer or inventory while it is
started or while it is stopped.
Prerequisites
To stop, start, or reset a single totalizer or inventory, the totalizer or inventory must be configured as a display
variable.
Procedure
To start or stop a single totalizer or inventory:
a) Wait or scroll until the totalizer or inventory appears on the display.
b) Choose Options.
c) Choose Start or Stop.
To start or stop all totalizers and inventories as a group:
a) Choose Menu > Operations > Totalizers.
b) Choose Start or Stop.
To reset a single totalizer or inventory:
a) Wait or scroll until the totalizer or inventory appears on the display.
b) Choose Options.
Transmitter operation Configuration and Use Manual