3. Check the wiring between the sensor and the transmitter.
4. Check for tube erosion, especially if the process fluid is abrasive.
5. Perform Smart Meter Verification.
15.7 Discrete Output problems
15.7.1 No Discrete Output
Cause
Output not powered
Wiring problem
Channel not configured for desired output
Channel not licensed
Circuit failure
Recommended actions
1. Check the power supply and power supply wiring.
2. Verify the output wiring.
3. Verify that the channel is wired and configured as a Discrete Output.
4. Purchase a license for the channel and upgrade the transmitter firmware.
5. Contact customer service.
15.7.2 Loop test failed
Cause
Output not powered
Power supply problem
Wiring problem
Circuit failure
Recommended actions
1. Check the power supply and power supply wiring.
2. Verify the output wiring.
3. Contact customer service.
15.7.3 Discrete Output readings reversed
Cause
Wiring problem
Configuration does not match wiring