Action Label
Display ProLink III Field Communicator
Totalizers
Start/stop all totalizers and
inventories
Start/stop all totalizers Start or Stop All Totalizers Start/Stop All Totals
Reset totalizer X Reset Total X Totalizer X Reset Total X
Reset all totalizers and
inventories
Reset All Totals Reset All Totals Reset All Totals
Concentration measurement
Increment CM matrix Increment Matrix Increment ED Curve Increment Curve
6.4 Configure totalizers and inventories
Display Menu > Configuration > Process Measurement > Totalizers & Inventories
ProLink III Device Tools > Totalizer Control > Totalizers
Field Communicator Configure > Manual Setup > Measurements > Optional Setup > Configure Totalizers
The transmitter provides seven configurable totalizers and seven configurable inventories. Each totalizer and
each inventory can be configured independently.
Totalizers track the process since the last totalizer reset. Inventories track the process since the last inventory
reset. Inventories are typically used to track the process across totalizer resets.
Tip
The default configurations cover the most typical uses of totalizers and inventories. You may not need to
change any configurations.
Prerequisites
Before configuring the totalizers and inventories, ensure that the process variables you plan to track are
available on the transmitter.
Procedure
1. Select the totalizer or inventory that you want to configure.
2. Set Totalizer Source or Inventory Source to the process variable that the totalizer or inventory will
track.
Option
Description
Mass flow The totalizer or inventory will track Mass Flow Rate and calculate total
mass since the last reset.
Volume flow The totalizer or inventory will track Volume Flow Rate and calculate total
volume since the last reset.
Gas standard volume
flow
The totalizer or inventory will track Gas Standard Volume Flow Rate and
calculate total volume since the last reset.
Configuration and Use Manual Configure advanced options for process measurement
MMI-20048166 April 2019
Configuration and Use Manual 73