Procedure
1. Select the event that you want to configure.
2. Assign a process variable to the event.
3. Specify Event Type.
Option Description
HI x > A
The event occurs when the value of the assigned process variable (x) is greater
than the setpoint (Setpoint A), endpoint not included.
LO x < A
The event occurs when the value of the assigned process variable (x) is less than
the setpoint (Setpoint A), endpoint not included.
IN A ≤ x ≤ B
The event occurs when the value of the assigned process variable (x) is in range,
that is, between Setpoint A and Setpoint B, endpoints included.
OUT x ≤ A or x ≥ B
The event occurs when the value of the assigned process variable (x) is out of
range, that is, less than Setpoint A or greater than Setpoint B, endpoints
included.
4. Set values for the required setpoints.
• For HI and LO events, set Setpoint A.
• For IN and OUT events, set Setpoint A and Setpoint B.
5. Optional: Configure a Discrete Output to switch states in response to the event status.
6. Optional: Specify the action or actions that the transmitter will perform when the event occurs.
• Using the display: Choose Menu > Configuration > Alert Setup > Enhanced Events, select any
enhanced event, and choose Assign Actions. Then assign the desired action to the desired
enhanced event.
• Using ProLink III: Choose Device Tools > Configuration > I/O > Inputs > Action Assignment. Then
assign the desired action to the desired enhanced event.
• Using the Field Communicator: Choose Configure > Alert Setup > Enhanced Events. Then assign
the desired action to the desired enhanced event.
Options for Enhanced Event Action
Action
Label
Display ProLink III Field Communicator
Standard
Start sensor zero Start Zero Calibration Start Sensor Zero Start Sensor Zero
Configure advanced options for process measurement Configuration and Use Manual
April 2019 MMI-20048166