8/20/2018 16:27 -0.20837 -0.20671 0.154502 0.157304 3.85E-05 1.291 …
8/21/2018 17:10 -0.11062 -0.11566 0.154641 0.157435 3.84E-05 1.288 …
8/22/2018 10:40 -0.15852 -0.16036 0.154512 0.157308 3.86E-05 1.284 …
8/25/2018 15:40 -0.00172 0.002301 0.154788 0.157599 4E-05 1.295 …
8/27/2018 23:16 0.132787 0.13684 0.155034 0.15785 4.08E-05 1.275 …
8/28/2018 11:16 0.04456 0.046158 0.154845 0.157653 3.99E-05 1.277 …
…
Note
The SMV log displays only in English.
14.4 Totalizer history and log
You can configure the transmitter to save totalizer and inventory values at a user-specified interval. You can
then generate a totalizer log. The totalizer log is a ASCII file.
Contents of totalizer log
The totalizer log contains one record for each logged totalizer or inventory value. Each record contains the
following information:
• Default totalizer or inventory name (user-specified names are not used)
• Value and measurement unit
• Timestamp
— Format: Military time
— Time and time zone: Transmitter clock
The totalizer log also contains a line item for each totalizer or inventory reset.
Totalizer history and power-cycles
If the transmitter is rebooted or power-cycled, totalizer history is not affected.
Totalizer history and configuration files
If you restore the factory configuration or upload a configuration file, totalizer history is not affected.
Totalizer log
================================================================================
Device UID: 22729F1F Device Tag: SUPPLY
Name Value Units Time Zone: GMT-7.00
================================================================================
Mass Fwd Total 61.74707 grams 9/12/2018 20:00
Mass Fwd Inv 61.74705 grams 9/12/2018 20:00
Mass Fwd Total 61.74707 grams 9/12/2018 21:00
Mass Fwd Inv 61.74705 grams 9/12/2018 21:00
Log files, history files, and service files Configuration and Use Manual
April 2019 MMI-20048166
150 Micro Motion 4200 Transmitters