4. Contact customer service.
15.2.5 [006] Transmitter Not Characterized Not Configured
Alert
Configuration Error
Cause
Combination of 020 and 021.
Recommended actions
1. Check the characterization. Specifically, verify the FCF and K1 values. Refer to Characterize the meter
(if required).
2. Contact customer service.
15.2.6 [008] Density Overrange
Alert
Extreme Primary Purpose Variable
Cause
The measured density is above 10 g/cm
3
.
Recommended actions
1. Check your process conditions against the values reported by the device.
2. Verify that the transmitter is configured correctly for the connected sensor.
3. Check for two-phase flow.
4. Verify all of the characterization or calibration parameters.
See the sensor tag or the calibration sheet for your meter.
5. Check the sensor coils for electrical shorts. If you find problems, replace the sensor.
6. Check the drive gain and the pickoff voltage.
7. Contact customer service.
15.2.7 [009] Transmitter Initializing/ Warming Up
Alert
Transmitter Initializing
Cause
The transmitter is in power-up mode.
Recommended actions
1. Allow the meter to complete its power-up sequence. The alert should clear automatically.
2. If the alert does not clear: