15.2.11 [013] Process too Noisy to perform Auto Zero
Alert
Noisy Process
Cause
Similar to 010.
Recommended actions
1. Remove or reduce sources of electromechanical noise, then retry. Sources of noise include:
Mechanical pumps
Pipe stress at sensor
Electrical interference
Vibration effects from nearby machinery
2. Reboot or power-cycle the flowmeter to see if the alert clears.
3. If appropriate, restore the factory zero to return the flowmeter to operation.
15.2.12 [014] Transmitter Failed Not Configured
Alert
Transmitter Initializing
Cause
Many possile causes.
Recommended actions
1. Reboot or power-cycle the transmitter to see if the alert clears.
2. Contact customer service.
15.2.13 [016] Sensor Temperature (RTD) Failure
Alert
Sensor Failed
Cause
The value computed for the resistance of the line RTD is outside limits.
Recommended actions
1. Check the wiring between the sensor and the transmitter.
a) Refer to the installation manual and ensure that the wiring has been performed according to
instructions. Obey all applicable safety messages.
b) Verify that the wires are making good contact with the terminals.
Configuration and Use Manual Troubleshooting