Procedure
1. Remove the transmitter-to-sensor wiring compartment cover to reveal the terminal
connections.
Figure 3-8: Removal of the transmitter-to-sensor wiring compartment cover
2. Feed the sensor wiring cable into the transmitter wiring compartment.
Figure 3-9: Sensor wiring feedthrough
3. Connect the sensor wires to the appropriate terminals.
Important
Terminate the four-wire cable drain wires only at the sensor/core processor end of
the cable. See the sensor installation manual for more detail. Do not connect the
four-wire cable drain wires to the ground screw located inside the junction box.
• See Figure 3-10 for four-wire terminal connections.
• See Figure 3-11 for nine-wire terminal connections.
Installation Manual Mounting and sensor wiring
MMI-20027478 April 2022
Installation Manual 17