Figure 3-10: Four-wire transmitter-to-sensor wiring connections
Figure 3-11: Nine-wire transmitter-to-sensor wiring connections
Note
Connect the four drain wires in the nine-wire cable to the ground screw located
inside the junction box.
4. Replace the transmitter-to-sensor wiring compartment cover and tighten the
screws to 14 in lbf (1.58 N m) to 15 in lbf (1.69 N m).
3.4 Ground the meter components
In 4-wire or 9-wire remote installations, the transmitter and sensor are grounded
separately.
Mounting and sensor wiring Installation Manual
April 2022 MMI-20027478