5 Wiring the power supply
You can install a user-supplied switch in the power supply line.
Important
For compliance with the Low Voltage Directive 2014/35/EU (European installations), verify
that the transmitter is located in close proximity to a switch.
Procedure
1. Remove the wiring access cover.
2. Open the Power warning flap to locate the power terminals.
Figure 5-1: Location of power supply wiring terminals and equipment ground
A
B
C
A. Power supply wiring terminals (+ and -)
B. Conduit connection for power supply wiring
C. Equipment ground
3. Connect the power supply wires.
For DC power: connect to terminals + and –.
For AC power: connect to terminals L/L1 (line) and N/L2 (neutral).
NOTICE
Do not use the conduit specified for power supply wires for input/output wiring (see
Figure 5-1) to avoid pinched wires when the cover is closed.
4. Tighten the two screws holding the power connector in place.
5. Ground the power supply using the equipment ground, also under the Power
warning flap.
Installation Manual Wiring the power supply
MMI-20027478 April 2022
Installation Manual 45