14.7 Status LED states
The status LED on the transmitter indicates whether or not alerts are active. If alerts are active, view the alert
list to identify the alerts, then take appropriate action to correct the alert condition.
Your meter has one or two status LEDs:
A status LED on the display (only if your meter has a display)
A status LED on the board, beneath the meter housing cover
CAUTION
If your meter is in a hazardous area, do not remove the meter housing cover. Use a different method to
determine meter status.
The status LEDs use colors and flashing to indicate device status.
Table 14-1: Transmitter status reported by status LED
LED state Description Recommendation
Green No alerts are active. Continue with configuration or process
measurement.
Yellow One or more low-severity alerts are active. A low-severity alert condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or
process measurement. If you choose, you can
identify and resolve the alert condition.
Flashing yellow Calibration in progress, or Known Density
Verification in progress.
The low-severity alert condition does not
affect measurement accuracy or output
behavior. You can continue with configuration
or process measurement. If you chose, you can
identify and resolve the alert condition.
Red One or more high-severity alerts are active. A high-severity alert condition affects
measurement accuracy and output behavior.
Resolve the alert condition before continuing.
Related information
View and acknowledge status alerts
14.8 Status alerts, causes, and recommendations
Table 14-2: Status alerts, causes, and recommendations
Alert number Alert title Possible causes Recommended actions
A001 EEPROM Error The transmitter has
detected a problem
communicating with the
sensor.
Cycle power to the meter.
Contact Micro Motion.
Troubleshooting Configuration and Use Manual
March 2021 MMI-20020954
128 Micro Motion Gas Specific Gravity Meters (SGM)

Table of Contents