Tip
Scroll Rate may not be available until you apply Auto Scroll.
9.2 Enable or disable the Acknowledge All Alerts display
command
ProLink III Device Tools → Configuration → Transmitter Display → Ack All
Field communicator Configure → Manual Setup → Display → Display Menus → Acknowledge All
You can configure whether or not the operator can use a single command to acknowledge all alerts from the
display.
Procedure
1. Ensure that the alert menu is accessible from the display.
To acknowledge alerts from the display, operators must have access to the alert menu.
2. Enable or disable Acknowledge All Alerts as desired.
Option Description
Enabled Default. Operators can use a single display command to acknowledge all alerts at once.
Disabled Operators cannot acknowledge all alerts at once. Each alert must be acknowledged
separately.
9.3 Configure security for the display menus
ProLink III
Device Tools → Configuration → Transmitter Display → Display Security
Field communicator Configure → Manual Setup → Display → Display Menus
You can control operator access to different sections of the display off-line menu. You can also configure a
passcode to control access.
Procedure
1. To control operator access to the maintenance section of the off-line menu, enable or disable Off-Line
Menu.
Option
Description
Enabled Default. Operator can access the maintenance section of the off-line menu. This access is
required for configuration and calibration, including Known Density Verification.
Disabled Operator cannot access the maintenance section of the off-line menu.
2. To control operator access to the alert menu, enable or disable Alert Menu.
Configuration and Use Manual Configure device options and preferences
MMI-20020954 March 2021
Configuration and Use Manual 79