7.8.3 Select the active concentration matrix using ProLink III
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose Device Tools Configuration Process Measurement Concentration Measurement.
2. Set Active Matrix to the matrix you want to use and click Change Matrix.
Configure process measurement using ProLink III Configuration and Use Manual