EasyManua.ls Logo

Emerson Micro Motion SGM - Page 74

Emerson Micro Motion SGM
190 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
c) Optional: Set Energy Flow Unit to the unit to be used to measure energy flow.
2. If you want to use fixed values for Percent CO, Percent CO2, Percent N2, and/or Percent H2:
a) Choose Configure Manual Setup Measurements Optional Setup Energy Content
Measurement Gas Composition.
b) Enter the fixed values, in % by volume.
3. If you want to poll for Percent CO, Percent CO2, Percent N2, and/or Percent H2:
a) Choose Configure Manual Setup Inputs/Outputs External Device Polling and click
External Device Polling.
b) Choose an unused polling slot.
c) Set Poll Control to Poll as Primary or Poll as Secondary.
d) Set External Device Tag to the HART tag of the external measurement device.
e) Set Polled Variable to the appropriate variable.
4. Optional: To configure Volume Flow (External) and Mass Flow (Calculated):
a) Choose Configure Manual Setup Inputs/Outputs External Device Polling.
b) Choose an unused polling slot.
c) Set Poll Control to Poll as Primary or Poll as Secondary.
d) Set External Device Tag to the HART tag of the external measurement device.
e) Set Polled Variable to Volume from Mag/Vortex Meter.
f) Choose Configure Manual Setup Measurements Optional Setup External Inputs
Configure External Inputs Volume.
g) Set Volume Flow Source to Enable.
h) Choose Configure Manual Setup Measurements Volume.
i) Set Volume Flow Rate Unit to the unit used by the external device.
j) Choose Configure Manual Setup Measurements Mass.
k) Set Mass Flow Rate Unit to the unit to be used for Mass Flow (Calculated).
5. Optional: To configure Mass Flow (External) and Volume Flow (Calculated):
a) Choose Configure Manual Setup Inputs/Outputs External Device Polling.
b) Choose an unused polling slot.
c) Set Poll Control to Poll as Primary or Poll as Secondary.
d) Set External Device Tag to the HART tag of the external measurement device.
e) Set Polled Variable to Mass Flow from Coriolis Meter.
f) Choose Configure Manual Setup Measurements Optional Setup External Inputs
Configure External Inputs Mass.
Configure process measurement using a field communicator Configuration and Use Manual
March 2021 MMI-20020954
74 Micro Motion Gas Specific Gravity Meters (SGM)

Table of Contents

Related product manuals