8. Enter the density of the secondary constituent of your process gas, in its pure form.
8.4.3 Select the active concentration matrix using a field
communicator
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose Configure → Manual Setup → Measurements → Optional Setup → Concentration
Measurement → CM Configuration.
2. Choose Configure → Manual Setup → Measurements → Optional Setup → Conc Measurement →
CM Configuration.
3. Set Active Matrix to the matrix you want to use.
Configure process measurement using a field communicator Configuration and Use Manual
March 2021 MMI-20020954
76 Micro Motion Gas Specific Gravity Meters (SGM)