Instruction Manual
IM-106-880C, OI
Section 7: Foundation
Fieldbus
April 2017
About AMS and DeltaV
Software
16
5
The following steps have to be performed to install a new device onto a DeltaV
TM
system:
From the start menu select DeltaV > Engineering > DeltaV Explorer.
Select/Expand “Library” (right below DeltaV_System).
Select “Fieldbus Devices”, using right mouse button. Click on “Fieldbus Devices”. This
will bring up a list of options.
From the list, select “Add Device Definition”... This should give you a “Browse for folder”
selection box. Browse to the directory that contains the 7 files needed to “register” a
new device with DeltaV. These file will consist of 3 *.dll files, *.sym, *.ffo, *.fhx and
*.reg file. The files probably will be on a floppy disk or a CD-ROM that accompanies your
device. On CD-ROMs delivered together with Emerson analyzers the files are located in
the directory \Fieldbus. Dependent on the existent system use the files of the
appropriate subdirectory.
After answering “yes“ to the first prompt, DeltaV will start the installation.
Figure 7-34 shows the “Exploring DeltaV“ screen for reference.