3.2.3 Install the sensor
Complete the following steps to install the Rosemount 499ACL-02 Sensor in the TCL
system.
Procedure
1. From inside the sample conditioning enclosure, thread the sensor cable or VP cable
through the gland on the upper left side.
Leave about one foot of cable inside the enclosure.
2. Wire the cable to the transmitter. Refer to Sensor wiring.
3. Remove the nut and adapter from the flow cell.
4. Slip the nut over the end of the sensor.
5. Thread the adapter onto the sensor. Hand-tighten only.
6. If you are using a VP cable, connect the cable to the sensor.
The connector and receptacle are keyed to ensure proper mating.
7. Once the key has slid into place, tighten the connection by turning the knurled ring
clockwise.
8. Remove the protective cap from the end of the sensor.
9. Insert the sensor in the flow cell. Hand tighten the nut.