See Figure 12-5.
Figure 12-5: Pump Cover
A. Pump cover
4. Using your thumb as shown in Figure 12-6, push the tubing fitting straight outward
until the fitting slides out of the socket.
Figure 12-6: Pushing the Tubing Fitting Out
5. Remove and discard the pump tubing.
6. Insert the new tubing one end at a time. Tongues on the sides of the gray fittings
at the end of the tube fit into receiving grooves in the pump casing. Push the fitting
in place until it clicks. Gently stretch the tubing over the rollers and insert the other
fitting into the receiving socket on the other side of the pump.
See Figure 12-7.
Figure 12-7: Inserting New Tubing
7. Replace the pump cover.
a) Place the cover on the pump casing.
See Figure 12-8.
Maintenance Reference Manual