12.3.7 Replacing the air pump diaphragm and check valves
Complete the following steps to replace the air pump diaphragm and check valves on your
total chlorine system.
WARNING
Hazardous voltage
Can cause severe injury or death.
Disconnect power before servicing.
Procedure
1. Disconnect power to the transmitter.
2. Refer to Figure 12-11. Disconnect the reagent and air injection tubes. Disconnect the
suction and discharge fitting by turning the Luer fitting in the direction shown in the
figure. Disconnect the air pump inlet tubing from the barbed fitting at the bottom of
the enclosure.
3. Remove the four screws (circled in Figure 12-12) holding the air pump access panel.
Pull out the pump and panel.
4. Disconnect the air inlet and outlet tubing from the air pump.
See Figure 12-13.
5. Remove the five screws (surrounded by squares in Figure 12-12) holding the air
pump to the access panel.
6. Pull the rubber base off the pump.
7. Using needle nose pliers, remove the air inlet fitting from the side of the air pump.
See Figure 12-15.
Figure 12-15: Inlet Fitting
8. Slide the pump assembly out of the air pump body.
See Figure 12-16.
Reference Manual Maintenance
00809-0200-3415 January2023
Rosemount TCL 56 63