SPECORD PLUS Installation and commissioning
19
5.1.3 Spatial requirements
Figure12SPECORD PLUS dimensions
The device is designed as a tabletop device and weighs approximately 22kg. A stable
table with a load capacity of 50kg is recommended as the installation location.
The device requires a minimum installation surface area of 680 x 720mm due to its di-
mensions and the required approx. 10cm clearance on the side of the electrical connec-
tions. Also, provide sufficient space for possible accessories and the control computer
The ventilation slits on the top and bottom sides of the device must not be covered.
Device dimensions
Device Width x length x height
SPECORD PLUS 590 x 690 x 290mm
APG autosampler (optional) 500 x 540 x 470mm
Temperature control device for Peltier-con-
trolled accessories (optional)
225 x 200 x 130mm
Heat exchanger for Peltier-controlled acces-
sories (optional)
225 x 200 x 175mm
The temperature control device and the heat exchanger can be stacked upon each other
to save space.
5.2 Installing the device
WARNING
Risk of electric shock
Check that the mains connection conditions match those indicated on the type plate on
the rear of the device.
} Remove the SPECORD PLUS and accessories from the transport packaging. Remove
the protective sleeve from the device.
Caution! Transport the SPECORD PLUS in upright position only and do not overturn
it when unpacking. Observe the markings on the transport packaging.
} Place the device and the PC on a suitable surface (laboratory desk).
} Allow the device to acclimatize until it has reached the ambient temperature.
} Remove the transport lock (→ "Removing and fastening the transport locks" 20).
} Connect one mains cable each to the SPECORD PLUS and the PC. Connect mains con-
nectors to sockets connected to the same power supply phase (if possible via socket
strips).
Do not connect the USB cable yet.