SPECORD PLUS Specifications
39
10 Specifications
10.1 Physical/optical performance data
10.1.1 Optical system
Spectrometer sys-
tem
Monochromator with imaging grating and aspherical, quartz-coated
optics
Internal holmium oxide filter
Detector Two silicon photo diodes
Peltier-controlled for SPECORD 210 and 250 PLUS
Sample position Additional measuring location for opaque samples
Large selection of accessories for optimized positioning of solid, liquid
and gaseous samples
Light source Combination of halogen and deuterium lamp
Lamp replacement adjustable between 300 and 450nm
10.1.2 SPECORD 50 PLUS
Optics Double-beam spectrophotometer with split
beam technology
Wavelength range 190 to1100 mm
Photometric display range -9to9 A
Photometric measuring range -3to3 A
Spectral bandwidth 1.4 nm
Spectral resolution capability for
toluol/hexane at 20–25°C
1.6to1.8
Wavelength accuracy
(deuterium line at 486nm)
±0.2 nm
Wavelength accuracy
(deuterium line at 656.1nm)
±0.1 nm
Wavelength accuracy
(holmium oxide filter at 360.9nm)*
± 0.5 nm
Wavelength reproducibility
(holmium oxide filter at 360.9nm)*
≤0.02 nm
Transmission zero point
(200–1000nm)
±0.05 %T
VIS photometric accuracy
(Hellma neutral glass filter F4 at 546nm)*
±0.003 A
UV photometric accuracy
(potassium dichromate)*
±0.010 A
Photometric accuracy
(potassium dichromate at 430nm)*
±0.010 A
Photometric reproducibility
(Hellma neutral glass filter F4 at 546nm)*
≤0.0005 A
Scattered light at 198nm (KCl)** ≤0.3 %T
Scattered light at 220nm (Nal) ≤0.03 %T
Scattered light at 240nm (Nal) ≤0.03 %T