SPECORD PLUS Specifications
43
10.2 General technical data of the SPECORD PLUS
Mass SPECORD 50 PLUS: 21.0kg
SPECORD 200 PLUS: 21.9kg
SPECORD 210 PLUS: 22.2kg
SPECORD 250 PLUS: 22.8kg
Floor space (width x depth) approx 900x690mm with PC
Dimensions (width x height x depth) 590 x 290 x 690mm
Sample chamber dimensions (width x height
x depth)
364 x 200 x 185mm
Mains voltage 85–264V/AC
Frequency 50 – 60Hz
Power consumption 200VA
Mains fuses 2 x T 3.15AH / 250V, type 19181 by Wick-
mann
Data connections 2 x RS-232 for accessories
1 x USB port for connecting an external PC
1 x RS-232 for connecting an external PC
Ambient conditions
Temperature during operation 15 to 35°C
Humidity during operation 90% at 30°C
Air pressure 0.7 to 1.06bar
Temperature and humidity during storage 15–55°C at 10–30% humidity, use dessi-
cant
Energy consumption
Energy consumption in different operating modes (SPECORD 210 PLUS)
Operating state Power
in [VA]
Power
in [W]
Energy consumption
per year in [kW]
3h / 5-day week
Energy consump-
tion per year in
[kW]
8h / 5-day week
Standby without
lamps
31.3 16.6 12.9 34.5
Standby, VIS lamp
only
69.1 40.4 31.5 84.0
Standby, VIS lamp
only
92.5 54 42.1 112.3
Standby, VIS and UV
lamp
128 77.8 60.7 161.8
Both lamps on, scan
only in UV
131.5 82.7 64.5 172.0
Both lamps on, scan
only in UV
130.2 79.6 62.1 165.6
The power values for the SPECORD 50 PLUS are lower by approx. 10VA / 7W in all op-
erating states.
The power values for the SPECORD 200 PLUS are lower by approx. 10VA / 7W in all
operating states.