SPECORD PLUS Care and maintenance
35
7.3 Replacing the fuses
} Switch off the SPECORD PLUS via the mains switch and disconnect the mains plug
from the device connection.
} Pull on the cover of the fuse holder and open it (→ "Connections and display ele-
ments on the right side of the device" 11).
} Replace defective mains fuses.
Only use the following fuses: T 3.15AH/250V, type 19181
} Close the fuse holder.
} Connect the mains cable to the mains connection on the SPECORD PLUS.
} Switch on the SPECORD PLUS.
ü The SPECORD PLUS is now ready to measure.