Cerabar M, Deltabar M, Deltapilot M Installation
Endress+Hauser 35
4.7.5 Seal for flange mounting
Distorted measurement results.
The seal is not allowed to press on the process isolating diaphragm as this could affect the
measurement result.
‣ Ensure that the seal is not touching the process isolating diaphragm.
A0017743
Fig. 19:
1 Process isolating diaphragm
2Seal
4.7.6 Wall and pipe mounting (optional)
Mounting bracket
Endress+Hauser offers a mounting bracket for installing on pipes or walls (for pipes from 1
1/4
" up to 2" diameter).
In the case of pipe mounting, the nuts on the bracket must be tightened uniformly with a
torque of at least 5 Nm (3.69 lbf ft).
A0028493
Maßeinheit mm (in)
122 (4.8)
52 (2.05)
86 (3.39)
70 (2.76)
140 (5.51)
158 (6.22)
175 (6.89)
ø6 (0.24)
ø42...60 (1.65...2.36)