Commissioning with the FF configuration program Cerabar M, Deltabar M, Deltapilot M
142 Endress+Hauser
9.4 Commissioning with device application
Commissioning comprises the following steps:
1. Function check ( ä 71)
2. Selecting the language, measuring mode and pressure unit
3. Position adjustment ( ä 144)
4. Configuring measurement:
Pressure measurement ( ä 145 ff)
Level measurement ( ä 146 ff)
Flow measurement (Deltabar M) (Deltabar) ( ä 155 ff)
9.4.1 Selecting the language, measuring mode and pressure unit
Language selection (Display Transducer Block)
Measuring mode selection (Pressure Transducer Block)
Parameter name Description
Language/
DISPLAY_LANGUAGE
Options
Index: 14
Data type: unsigned8
Access: wr for Auto, OOS
Select language.
Options:
•English
Possibly another language (as selected when ordering the device)
One further language (language of the manufacturing plant)
Factory setting:
English
Parameter name Description
Measuring mode/
OPERATING_MODE
Index: 42
Data type: unsigned8
Access: OOS
Select the measuring mode.
The operating menu is structured differently depending on the measuring mode
selected.
If the measuring mode is changed, no conversion takes place. If necessary, the
device has to be recalibrated after the measuring mode has been changed.
Options:
Pressure
•Level
•Flow (Deltabar)
Factory setting:
Pressure