EasyManua.ls Logo

Endress+Hauser Cerabar M - Installing Cerabar M

Endress+Hauser Cerabar M
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation Cerabar M, Deltabar M, Deltapilot M
14 Endress+Hauser
4.5 Installing Cerabar M
Due to the orientation of the Cerabar M, there may be a shift in the zero point, i.e. when
the container is empty or partially full, the measured value does not display zero. You can
correct this zero point shift ä 47, Section "Function of the operating elements" or
ä 73, Section 8.3 "Pos. zero adjust".
For PMP55, please refer to Section 4.5.2 "Installation instructions for devices with
diaphragm seals – PMP55", ä 17.
Endress+Hauser offers a mounting bracket for installing on pipes or walls.
ä 21, Section 4.5.5 "Wall and pipe mounting (optional)".
4.5.1 Installation instructions for devices without diaphragm seals –
PMP51, PMC51
NOTICE
Damage to the device!
If a heated Cerabar M is cooled during the cleaning process (e.g. by cold water), a vacuum
develops for a short time, whereby moisture can penetrate the sensor through the pressure
compensation (1).
If this is the case, mount the Cerabar M with the pressure compensation (1) pointing
downwards.
Keep the pressure compensation and GORE-TEX
®
filter (1) free from contamination.
Cerabar M transmitters without diaphragm seals are mounted as per the norms for a
manometer (DIN EN 837-2). We recommend the use of shutoff devices and siphons. The
orientation depends on the measuring application.
Do not clean or touch process isolating diaphragms with hard or pointed objects.
The device must be installed as follows in order to comply with the cleanability
requirements of the ASME-BPE (Part SD Cleanability):
1
1
1

Table of Contents

Related product manuals