Deltabar FMD71, FMD72
Endress+Hauser 39
FMD71 process connections,
flush-mounted process
isolating diaphragm
Hygienic connections
A
ø42 (1.65)
ø64 (2.52)
ø61 (2.4)
ø57 (2.24)
ø52 (2.05)
13 (0.51)
20 (0.79)
72 (2.83)
A0026375
B
ø42 (1.65)
ø52 (2.05)
ø77 (3.03)
ø91 (3.58)
ø84 (3.31)
72 (2.83)
A0026376
A0026376
C
ø42 (1.65)
ø75 (2.95)
ø84 (3.31)
ø52 (2.05)
12 (0.47)
20 (0.79)
72 (2.83)
A0026377
D
ø38 (1.5)
ø66 (2.6)
ø52 (2.05)
ø60 (2.36)
12 (0.47)
42 (1.65)
49 (1.93)
A0026378
Engineering unit mm (in)
Item Description Material Weight Option
1)
kg (lbs)
A
2)
Tri-Clamp ISO 2852 DN 40 – DN 51 (2"), DIN 32676DN 50, EHEDG, 3A (CRN) AISI 316L (1.4435) 0.7 (1.54) TDJ
3)
B Tri-Clamp ISO 2852 DN 76.1 (3"), EHEDG, 3A, with FDA seal 0.9 (1.98) TFJ
3)
C
2)
Varivent Type N for pipes 40 – 162, PN 40, EHEDG, 3A (CRN) 1 (2.21) TRJ
D Varivent Type F for pipes DN25-32 PN 40, 316L, EHEDG, 3A, with FDA seal 0.46 (1) TQJ
1) Product Configurator order code for "Process connection"
2) Roughness of wetted surfaces R
a
< 0.76 µm (30 µin) as standard. Surface quality R
a
< 0.38 µm (15 µin) electropolished (wetted) available on
request.
3) Endress+Hauser supplies these slotted nuts in stainless steel AISI 304 (DIN/EN material number 1.4301) or in AISI 304L (DIN/EN material
number 1.4307).