EasyManua.ls Logo

Endress+Hauser Deltabar FMD72 - Device Features; Product Design

Endress+Hauser Deltabar FMD72
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Deltabar FMD71, FMD72
8 Endress+Hauser
Device features
FMD71 FMD72
Field of
application
Level
Differential pressure
Process
connections
Thread
EN and ASME flanges
Flush-mounted hygienic connections
Thread
EN and ASME flanges
Flush-mounted hygienic connections
Measuring ranges
(differential
pressure)
from
–100 to +100 mbar (–1.5 to +1.5 psi)
to –1 to +40 bar (–15 to +600 psi)
from –400 to +400 mbar (–6 to +6 psi)
to –1 to +40 bar (–15 to +600 psi)
MWP
1)
to 60 bar (900 psi) to 160 bar (2 400 psi)
Process
temperature
limits
Thread / flange connections:
–25 to +125 °C (–13 to +257 °F)/
150 °C (302 °F)
Hygienic connections:
–25 to +130 °C (–13 to +266 °F) ,
150 °C (302 °F) for max. 1 hour
Flush-mounted thread G 1 A, G 1½, G 2,
MNPT / flange connections:
–40 to +100 °C (–40 to +212 °F)/150 °C (302 °F)
Thread with internal process isolating
diaphragm: –40 to +125 °C (–40 to +257 °F)/
150 °C (302 °F)
Hygienic connections:
–40 to +130 °C (–40 to +266 °F) ,
150 °C (302 °F) for max. 1 hour
Higher process temperature limits available on
request (up to +250 °C (480 °F))
Ambient
temperature
range
–40 to +80 °C (–40 to +176 °F)
Reference
accuracy of the
individual sensors
Up to ±0.075 % of the set span
PLATINUM version: up to ±0.05 % of the set span
Supply voltage Version for non-hazardous areas: 12 (13) to 45 V DC
Ex ia: 12 (13) to 30 V DC
Output 4 to 20 mA with superimposed HART protocol
Options NACE-compliant materials
Software pre-configurations
1) MWP: depends on the lowest-rated element, with regard to pressure, of the selected components