Deltabar PMD75B PROFINET with Ethernet-APL Diagnostics and troubleshooting
Endress+Hauser 73
11.8.3 Resetting the device via keys on the electronic insert
Resetting the password
Display
onoff
15 s
2×
I I I I
I
A0050210
10 Sequence - resetting the password
Deleting/resetting the password
1. Press operating key I three times.
The Reset Password function is started, the LED flashes.
2. Press operating key I once within 15 s.
The password is reset, the LED flashes briefly.
If operating key I is not pressed within 15 s, the action is canceled and the LED is no longer
lit.
Resetting the device to the factory setting
A0050015
11 Sequence - factory setting
Resetting the device to the factory setting
‣
Press operating key I for at least 12 s.
Device data are reset to the factory setting, the LED flashes briefly.
11.9 Firmware history
The firmware version can explicitly be ordered via the product structure. This makes it
possible to ensure the compatibility of the firmware version with an existing or
planned system integration.