Appendix Deltapilot S FMB70
34 Endress+Hauser
8.2.1 Pressure device parameter configuration form
Operation via: Handheld terminal FieldCare/DeviceCare On site
Device designation: Serial number:
Measuring
point:
Upper range limit (URL sensor):
Parameter name Direct
access
Menu path:
Expert
Factory setting
1)
Permitted
settings
Set
point
Actual
value
Tested
Digit set 840 → Display 012345
67890.-
Calib. offset 319 → Position
adjustment
0.0
2)
Measuring mode 389 Pressure Pressure
Set LRV 245 → Quick Setup /
Basic Setup
0.0 or as per order
2)
Set URV 246 URL sensor or as per order
2)
Damping value 247 2 s 0 to 999 s
Press. eng. unit 060 → Basic setup mbar/bar
or as per order
3)
All
except "user unit"
Current charact. 695 → Output Linear Linear
Output fail mode 388 Max. alarm Max. alarm
Min. alarm
Alt. curr. output 597 Normal Normal
Set max. alarm 342 22 mA 21 to 23 mA
Set min. current 343 3.8 mA 3.8 mA or 4 mA
Simulation 413 → Simulation None None
Ack. alarm mode 401 → Messages Off Off/On
Error No. Select alarm
type
The following messages
must be set to "Alarm":
595/600
115 Sensor over pressure Warning Alarm o. warning
120 Sensor low pressure Warning Alarm o. warning
715 Sensor over
temperature
Warning Alarm o. warning
717 Transmitter over
temp.
Warning Alarm o. warning
718 Transmitter under
temp.
Warning Alarm o. warning
720 Sensor under
temperature
Warning Alarm o. warning
726 Sens. temp. error
overrange
Warning/alarm
4)
Alarm
727 Sens. pres. error
overrange
5)
Warning Alarm o. warning
620 Current output out
of range
Warning Alarm o. warning
Alarm delay 336 0.0 s 0.0 s
Alarm display time 480 0.0 s 0.0 s