Operation Liquiline CM442/CM444/CM448
92 Endress+Hauser
Path: Menu/Setup/Additional functions/Limit switches/Limit switches 1 to 8
Function Options Info
Cleaning program Options
•None
Cleaning 1
Cleaning 2
Cleaning 3
Cleaning 4
Factory setting
None
Use this function to choose the cleaning instance which
should be started when the limit switch is active.
Function Options
•On
•Off
Factory setting
Off
Activating/deactivating the limit switch
Operating mode Options
Above limit check
Below limit check
In range check
Out of range check
•Change rate
Factory setting
Above limit check
Type of limit value monitoring:
Limit value overshoot or undershoot å 86
Measured value within or outside a range å 87
•aChange rate å 89
Limit value Settings
Depends on the measured
value
Operating mode="Above limit check" or "Below limit
check"
a0018080
Fig. 86: Exceeding (A) and undershooting (B) a limit value (without hysteresis and switch-on delay)
1 Limit value
2 Alarm range
t
1
,t
3
,t
5
No action
t
2
,t
4
An event is generated
If the measured values (MV) are increasing, the relay contact is closed when the on-value is exceeded ("Limit value" +
"Hysteresis") and the start delay has elapsed ("Start delay").
If the measured values are decreasing, the relay contact is reset when the off-value is undershot ("Limit value" -
"Hysteresis") and following the release delay ("Switch off delay").