EasyManua.ls Logo

Endress+Hauser Micropilot FMR56 - Page 26

Endress+Hauser Micropilot FMR56
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR56, FMR57
26 Endress+Hauser
Supply voltage
An external power supply is necessary.
Various power supply units can be ordered from Endress+Hauser: see "Accessories" section
2-wire, 4-20mA HART, passive
"Power supply,
output"
1)
"Approval"
2)
Terminal voltage U
at device
Maximum load R, depending on the supply voltage U
0
of the power supply unit
A: 2-wire;
4-20mA HART
Non-Ex
Ex nA
Ex ic
CSA GP
10.4 to 35 V
3)
4) 5)
R [ ]?
U
0
[V]
10
10.4 21.4
20 30 35
0
500
A0017140
Ex ia / IS 10.4 to 30 V
3) 4) 5)
Ex d(ia) / XP
Ex ic(ia)
Ex nA(ia)
Ex ta / DIP
13 to 35 V
5) 6)
U
0
[V]
10
13 24
20 30 35
0
500
R [ ]Ω
A0034771
Ex ia + Ex d(ia) / IS +
XP
13 to 30 V
5) 6)
1) Feature 020 in the product structure
2) Feature 010 in the product structure
3) At ambient temperatures T
a
≤ -20 °C, a terminal voltage U ≥ 15 V is required to start the device with the minimum failure current (3.6 mA). The
start-up current can be configured. If the device is operated with a fixed current I ≥ 5.5 mA (HART Multidrop mode), a voltage U ≥ 10.4 V suffices
in the entire ambient temperature range.
4) A voltage U ≥ 12.5 V is required in the current simulation mode.
5) If the Bluetooth module is used, the minimum supply voltage increases by 3 V.
6) At ambient temperatures T
a
≤ -20 °C, a terminal voltage U ≥ 16 V is required to start the device with the minimum failure current (3.6 mA).
"Power supply,
output"
1)
"Approval"
2)
Terminal
voltage U at
device
Maximum load R, depending on the supply voltage U
0
of the power supply
unit
B: 2-wire; 4-20 mA
HART, switch
output
Non-Ex
Ex nA
Ex nA(ia)
Ex ic
Ex ic(ia)
Ex d(ia) / XP
Ex ta / DIP
CSA GP
13 to 35 V
3) 4)
U
0
[V]
10
13 24
20 30 35
0
500
R [ ]Ω
A0034771
Ex ia / IS
Ex ia + Ex d(ia) / IS + XP
13 to 30 V
3) 4)
1) Feature 020 in the product structure
2) Feature 010 in the product structure
3) At ambient temperatures T
a
≤ -30 °C, a terminal voltage U ≥ 16 V is required to start the device with the minimum failure current (3.6 mA).
4) If the Bluetooth module is used, the minimum supply voltage increases by 3 V.

Other manuals for Endress+Hauser Micropilot FMR56

Related product manuals