36
Nivotester FTR525
Endress+Hauser
Commissioning
8�4�2�12 Switch-ondelayswitchingoutput1
Application
2000
→
Switchingoutputs
2400
→
Switchingoutput1
2410
→
Switch-ondelayoutput1
2414
delay is enabled, the measured value must be above the switch-on point (function 2412)
for at least the set time before the switching output is activated.
Example: Switchdelaysoflimitdetection
The following diagram shows the behavior of the switching output with activated
2415), depending on additional adjustments
(assignment outputs in function group 2200, switching outputs in function group 2400).
T
ON
T
ON
T
OFF
T
OFF
26 Switch-onandswitch-offdelays
8�4�2�13 Switch-odelayswitchingoutput1
Application
2000
→
Switchingoutputs
2400
→
Switchingoutput1
2410
→
Switch-odelayoutput1
2415
T
OFF
for switching output 1. While
2413) for at least the set time before the switching output is deactivated (see example
function 2414).
8.4.2.14 Switchingfunctionswitchingoutput2
Application
2000
→
Switchingoutputs
2400
→
Switchingoutput2
2420
→
Switchingfunctionoutput2
2421
function 2411
Function group 2420 for switching output 2 is only available for device versions with two
switching outputs.
8.4.2.15 Switch-onpointswitchingoutput2
Application
2000
→
Switchingoutputs
2400
→
Switchingoutput2
2420
→
Switch-onpointoutput2
2422
In this function the switching point of switching output 2, below which the switching
(function 241235).