C
Courtney HawkinsAug 13, 2025
What to do if local display is dark and there are no output signals in Endress+Hauser Measuring Instruments?
- MMorgan TaylorAug 13, 2025
If the local display is dark and there are no output signals, there might be several reasons: * The supply voltage may not match the value indicated on the nameplate. In this case, apply the correct supply voltage. * The polarity of the supply voltage could be wrong; correct the polarity. * There might be no contact between connecting cables and terminals. Check the connection of the cables and correct if necessary. * Terminals may not be correctly plugged into the I/O or main electronics module; check the terminals. * The I/O or main electronics module could be defective; in this case, order spare part ? ? 182.