Proline Promag H 300
Endress+Hauser 41
If operating outdoors:
Install the measuring device in a shady location.
Avoid direct sunlight, particularly in warm climatic regions.
Avoid direct exposure to weather conditions.
You can order a weather protection cover from Endress+Hauser : →  91
Storage temperature
–50 to +80 °C (–58 to +176 °F)
Protect the measuring device against direct sunlight during storage in order to avoid unacceptably
high surface temperatures.
Select a storage location where moisture cannot collect in the measuring device as fungus or
bacteria infestation can damage the liner.
If protection caps or protective covers are mounted these should never be removed before
installing the measuring device.
Degree of protection
Transmitter and sensor
As standard: IP66/67, type 4X enclosure
With the order code for "Sensor options", option CM: IP69K can also be ordered
When housing is open: IP20, type 1 enclosure
Display module: IP20, type 1 enclosure
External WLAN antenna
IP67
Vibration resistance
Vibration, sinusoidal according to IEC 60068-2-6
2 to 8.4 Hz, 3.5 mm peak
8.4 to 2 000 Hz, 1 g peak
Vibration broad-band random, according to IEC 60068-2-64
10 to 200 Hz, 0.003 g
2
/Hz
200 to 2 000 Hz, 0.001 g
2
/Hz
Total: 1.54 g rms
Shock resistance
Shock, half-sine according to IEC 60068-2-27
6 ms 30 g
Impact resistance
Rough handling shocks according to IEC 60068-2-31
Mechanical load
Protect the transmitter housing against mechanical effects, such as shock or impact.
Never use the transmitter housing as a ladder or climbing aid.
Interior cleaning
Cleaning in place (CIP)
Sterilization in place (SIP)
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
For details, refer to the Declaration of Conformity.
Process
Medium temperature range
–20 to +150 °C (–4 to +302 °F)