EasyManua.ls Logo

Endress+Hauser Proline Promag H 300 - Sanitary compatibility; Functional safety; HART certification; FOUNDATION Fieldbus certification

Endress+Hauser Proline Promag H 300
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promag H 300
Endress+Hauser 89
Sanitary compatibility
3A approval and EHEDG-certified
Seals → FDA-compliant (apart from Kalrez seals)
Functional safety
The measuring device can be used for flow monitoring systems (min., max., range) up to SIL 2
(single-channel architecture; order code for "Additional approval", option LA) and SIL 3
(multichannel architecture with homogeneous redundancy) and is independently evaluated and
certified by the TÜV in accordance with IEC 61508.
The following types of monitoring in safety equipment are possible:
Volume flow
Functional Safety Manual with information on the SIL device →  94
HART certification HART interface
The measuring device is certified and registered by the FieldComm Group. The measuring system
meets all the requirements of the following specifications:
Certified according to HART 7
The device can also be operated with certified devices of other manufacturers (interoperability)
FOUNDATION Fieldbus
certification
FOUNDATION Fieldbus interface
The measuring device is certified and registered by the FieldComm Group. The measuring system
meets all the requirements of the following specifications:
Certified in accordance with FOUNDATION Fieldbus H1
Interoperability Test Kit (ITK), revision version 6.1.2 (certificate available on request)
Physical Layer Conformance Test
The device can also be operated with certified devices of other manufacturers (interoperability)
Certification PROFIBUS PROFIBUS interface
The measuring device is certified and registered by the PROFIBUS User Organization (PNO). The
measuring system meets all the requirements of the following specifications:
Certified in accordance with PROFIBUS PA Profile 3.02
The device can also be operated with certified devices of other manufacturers (interoperability)
Modbus RS485 certification
The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the
"MODBUS/TCP Conformance Test Policy, Version 2.0". The measuring device has successfully passed
all the test procedures carried out.
Radio approval
Europe:
RED 2014/53/EU
United States of America:
CFR Title 47, FCC Part 15.247
Canada:
RSS-247 Issue 1
Japan:
Article 2 clause 1 item 19
Additional country-specific approvals on request.
Other standards and
guidelines
EN 60529
Degrees of protection provided by enclosures (IP code)
EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use -
general requirements
IEC/EN 61326
Emission in accordance with Class A requirements. Electromagnetic compatibility (EMC
requirements).
NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment
NAMUR NE 32
Data retention in the event of a power failure in field and control instruments with
microprocessors

Table of Contents