48 Endress+Hauser
Process connection: flange according to DIN 11864-2
0
5
10
15
20
25
30
[bar][psi]
-40
-20
0 20
40 60 80 100 140 180 [°C]
360 [°F]
0-40 100 200 300
200
100
400
300
0
PN25 DN40 1 ½"): (
120 160
PN16 DN50...100 2...4"): (
PN10 DN125...150 5...6"): (
A0028945-EN
 33 Process connection material: stainless steel, 1.4404 (F316L)
Pressure tightness
Liner: PFA
Nominal diameter Limit values for absolute pressure in [mbar] ([psi]) for fluid temperatures:
[mm] [in] +25 °C
(+77 °F)
+80 °C
(+176 °F)
+100 °C
(+212 °F)
+130 °C
(+266 °F)
+150 °C
(+302 °F)
2 to 150 ¹⁄₁₂ to 6 0 (0) 0 (0) 0 (0) 0 (0) 0 (0)
Flow limit
The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The
optimum velocity of flow is between 2 to 3 m/s (6.56 to 9.84 ft/s). Also match the velocity of flow
(v) to the physical properties of the fluid:
v < 2 m/s (6.56 ft/s): for low conductivity values
v > 2 m/s (6.56 ft/s): for fluids producing buildup (e.g. milk with a high fat content)
A necessary increase in the flow velocity can be achieved by reducing the sensor nominal
diameter.
For an overview of the full scale values for the measuring range, see the "Measuring range"
section →  8
Pressure loss
No pressure loss occurs as of nominal diameter DN 8 (3/8") if the sensor is installed in a pipe with
the same nominal diameter.
Pressure losses for configurations incorporating adapters according to DIN EN 545 →  39
System pressure
A0028777
Never install the sensor on the pump suction side in order to avoid the risk of low pressure, and thus
damage to the liner.
Furthermore, install pulse dampers if reciprocating, diaphragm or peristaltic pumps are used.
For information on the liner's resistance to partial vacuum
For information on the shock resistance of the measuring system
For information on the vibration resistance of the measuring system
Vibrations
In the event of very strong vibrations, the pipe and sensor must be supported and fixed.
For information on the shock resistance of the measuring system
For information on the vibration resistance of the measuring system